HST 216/343/640/1000 BDS Dual Dust & Sand Chamber
Description
Nviro Solutions Dust Chambers are made for the simulation of Dust and Sand conditions in a controlled atmosphere.
Standard
MIL STD 810F - Method 510.4.Para 4.2.2, Procedure 1 - Blowing Dust
MIL STD 810F - Method 510.4.Para 4.2.3, Procedure II - Blowing Sand
Features
- Interior of chamber made of high quality stainless steel (SUS#304). Exterior of chamber is made of solid baked painted Steel plate. Interior of chamber treated with high temperature protection material. Chamber body connected to ground for ant-static protection.
- The Airflow Cycle System uses Baked Painted Steel Plate with over temperature protection.
- High pressure air circulating fan for wind speed 1.0 ~ 29.0 m/s attached with a wind speed adjustor.
- Electric heater with SSR & PID control for constant temperature.
- Dust Supplementary Device used for automatically adjusting the dust volume based on set values.
- Automatic Dry Air Handling System
- Chamber is well sealed and sealing must be maintained at all times to meet the environmental requirements of Dust/Sand and Powder exhaust.
- Control System consists of Temperature Control Unit, Humidity Control Unit, Wind Speed Control Unit and Concentration Control Unit.
- Control Panel consists of LED Controller (Other Options available) and DATA Collection.
- Uses electronic concentration sensors to automatically adjust and set dust and sand consistency levels.
Specifications
Specification | Specification in US | Specification in Metric |
---|---|---|
Temperature Range | 73.4°F to 158°F (tolerance ?± 2°F) | 23°C to 70°C (tolerance ?± 2°C) |
Humidity Range | <30% RH | <30% RH |
Air Speed | 1.0 to 10.0 m/s (Procedure I) 18.0 to 29.0 m/s (Procedure II) | 1.0 to 10.0 m/s (Procedure I) 18.0 to 29.0 m/s (Procedure II) |
Dust & Sand Consistency | 3.0 to 17.6 g/m³ (10.7 ± 7 g/m³) (Procedure I) 0.18 + 0.2 g/m³; 1.1 ± 0.3 g/m³; 2.2 ± 0.5 g/m³ (Procedure II) | 3.0 to 17.6 g/m³ (10.7 ± 7 g/m³) (Procedure I) 0.18 + 0.2 g/m³; 1.1 ± 0.3 g/m³; 2.2 ± 0.5 g/m³ (Procedure II) |
Interior Dimension | 23.6W x 23.6H x 23.6D IN | 600W x 600H x 600D mm |
Exterior Dimension | 165W x 86.6H x 78.7D IN (approx.) | 4200W x 2200H x 2000D mm (approx.) |
Door | 31.5W x 27.6H IN (Single Door) | 800W x 700H mm (Single Door) |
Observation Window | 11.8W x 11.8H IN | 300W x 300H mm |
Power Sources | 380/440/480VAC/50/60Hz | 380/440/480VAC/50/60Hz |